Aitek Display Offline Resolution for Tachtrol 30 Tachtrol 10 & Tachtrol Plus
“Display Is Offline” may be displayed on the front panel of a TACHTROL tachometer or remote display. In most cases this is a simple fix. Displays communicate to the tachometer through a local network, created by the tachometer, regardless of if they are part of the instrument (TACHTROL 10 & TACHTROL 30) or as a remote display (TACHTROL plus). Each display must have a distinct address to be properly orchestrated on the network and there can be up to 8, including the TACHTROL instrument display, on any local tachometer network.
- From the front panel press F2 > Security
- On the drop down menu verify Display Address is not “0”. It can be any number between 1 & 8. If there will be multiple remote displays connected it is best to use 1 for the tachometer.
- If the display address = 0 it must be changed.
- Use the down arrow (4) to navigate to Display Address.
- Press Enter to toggle the address number. Set to the desired number. If there are additional TACHTOL plus units connected, each display must be a different number.
- Press F1 to return to the main screen