Tuesday, June 12, 2012

Purpose of a Vent Hole in Pressure Sensors

Author: Michele Beyer (WIKA)
Posted: Knowledge

In so-called “gauge sensors” pressure is always measured relative to the currently prevailing ambient pressure. However, the ambient pressure is by no means constant. It is location-and height-dependent and changes with the weather. This means that for a “gauge” pressure sensor to operate, the (changing) ambient pressure must also be available inside the instrument. This means that the ambient pressure must be allowed to enter the instrument, i.e., the instrument must be “vented”. If an instrument of this type were to remain unvented, the change in weather and location, relative to the calibration performed by the manufacturer, an additional error up to several per cent would be introduced. This effect becomes of course less and less important with increasing pressure range of the sensor, compared with the actual change in ambient pressure.

By the way: in unvented instruments, the temperature influence adds another effect. Expansion of the air enclosed in the instrument as a result of a change in temperature causes a preliminary pressure, which also appears as an additional measuring error.

Conclusion: Pressure sensors used for measuring gauge pressure must be “vented”.

Typically, this is done via a controlled opening, a so-called vent hole. This vent hole must, of course, be carefully protected from the entry of dirt, dust and moisture. This can be supported by design measures (location, membranes, labyrinths, etc.) or must be ensured by the user on site by carefully controlling the mounting situation.

If you would like more information about WIKA pressure sensors or other WIKA products please contact Forberg Scientific customer service.
Toll Free: 855-288-5330
Email: mechanicalsales@forberg.com

Friday, January 13, 2012

Active & passive temperature compensation of pressure sensors

Author: Dominik Lorenz
Posted: Knowledge

How does an active or passive temperature compensation of pressure transmitters actually work? High-quality pressure transmitters, especially those used in precision critical applications, are almost always provided with an individual temperature compensation. But what is actually the difference between an active and a passive temperature compensation of these sensors?

Passive temperature compensation:
Sections of the characteristic accuracy curve of the pressure sensor are measured at different temperatures during the manufacturing process. Then, the previously determined temperature errors are compensated by passive elements (resistors) within the electronics of the sensor or by corrections of specifically designed resistance structures directly on the sensor element itself (e.g. by laser-trimming). The (passive) resistor elements used have an almost linear temperature behaviour, it is, however, only possible to compensate 1st order errors. Temperature errors of higher order, i.e. strong bending of the characteristic curve under temperature, can not be compensated.

Active temperature compensation:
Here too, the characteristic curve of the pressure transmitters is measured at different temperatures during the manufacturing process. However, the pressure transmitter has an additional integrated temperature sensor which constantly measures the temperature of the sensor and transfers it to the pressure transmitter’s signal processing. In practice, two methods of active temperature compensation are common: the first method compensates by means of a limited number of samples, , i.e. discrete correction values, between which interpolation takes place. The second method uses the electronics of the transmitter and a higher-order equation resulting from the regression of the acquired measurement values in order to compensate then the expected error.

During operation, this signal processing makes it possible to automatically, i.e. “actively”, compensate the pressure transmitter‘s temperature error using the calculated correction factors within a specified temperature range 50-140°F (e.g. 10-60°C).

The most commonly used method to minimise temperature errors of pressure sensors is a passive temperature compensation. This is the traditional method which is widely used. However, active temperature compensation is the top class of possible compensation methods. WIKA has constantly improved and refined this technology in recent years.

The pressure transmitters of WIKA using active temperature compensation therefore feature a temperature error which is almost zero in their specified temperature range.

If you would like to get more information about WIKA pressure transmitters or other WIKA products please contace Forberg Scientific, Inc. customer service.
Toll Free: 1-855-288-5330
Email: mechanicalsales@forberg.com

Friday, September 23, 2011

Diaphragm Seal & Chemical Seal Operating Principle

The drawing below illustrates the operating principle of a diaphragm seal assembly. A pressure measurementinstrument such as a conventional pressure gauge or electronic pressure transmitter is either mounted directly to the diaphragm seal or attached to the seal by means of a capillary or cooling element.
A diaphragm within the diaphragm seal separates the gauge / transmitter from the process medium. Any part of the diaphragm seal (i.e., diaphragm, lower housing, gaskets) which will be exposed to the process medium is selected from materials resistant to pressure, temperature and possible chemical attack by the process medium.

The diaphragm seal is also filled with a transmitting fluid or system fill fluid. Any pressure applied by the process medium to the seal diaphragm is hydraulically transmitted to the pressure element of the gauge / switch / transmitter thus generating a pressure reading.

If you would like additional information about WIKA Diaphragm Seals or other WIKA Products please contact Forberg Scientific, Inc.
Toll Free: 855-288-5330
Email: mechanicalsales@forberg.com

Wednesday, July 13, 2011

Safety Critical Protection Systems: Implementing Pressure Sensor Signal Redundancy to Lower Costs

Written By: Jürgen Reiser, Product Manager, Electronic Pressure Measurement, WIKA Alexander Wiegand GmbH & Co. KG, Germany and Martin Armson, Director of Electronic Pressure Measurement, WIKA Instrument Corporation, USA

In today’s world, safe working conditions are increasingly mandated by legislation and voluntary codes of practice. Automated load monitoring systems using pressure transmitters are becoming instrumental in creating safe working environments. They play an important part in making certain that machine control systems used in utility vehicles including cranes, lifting platforms and fork-lift trucks perform with precision, accuracy, and most importantly reliability. ‘Failsafe’ is the keyword that drives system designers to be ever more creative. System redundancy is a design concept that helps meet these regulatory requirements. Pressure transmitter based systems must guarantee that systematic failures within hydraulic controls are eliminated and random error functions are identified so that all moving parts are shut down if a dangerous condition exists.

In typical load and load-moment monitoring applications, pressure is measured at the hydraulic cylinder of the boom, the hydraulic pump of the traction drive, or at the pressure reservoir of the brake system. Any malfunction of a pressure transmitter may lead to a critical condition of the system. A malfunction may be caused by extreme environmental conditions (EMC, humidity, shock and vibration, or physical damage). It can also be caused by abnormal operating conditions including an overload of the pressure sensor caused by pressure spikes. Erratic software failures in the subsystem program or a defective electronic component may also be a source of an incorrect pressure transmitter and/or system output signal.


Machine designers take many preventative steps to provide a continuously safe operating condition. One easy, practical, and common solution is to use two pressure sensors at the same measurement point. A redundancy check of the two output signals is made in the machine logic control circuit to detect any unexpected deviation in pressure transmitter readings.

The disadvantages of this approach are that an additional pressure port with separate sealing is required, as are additional electrical connectors with mating connectors. A larger number of electrical components are required to use two individual pressure sensors. From a safety standpoint, this increased complexity presents additional opportunities for failure.

The WIKA Answer: ‘2 in 1‘ Pressure Sensor MHS-1

WIKA’s MHS-1 pressure transmitter offers the first integrated dual pressure sensors, manufactured specifically for safety- critical and safety-related applications in mobile hydraulics. Dual (2 channel) pressure measurement, separate electrical signal processing and two independent 4-20mA output signals enable control-side redundant monitoring for pressure measurement points.

It is now possible for Channel 1 to be a pressure sensor with measuring range 0-3000 PSI with a 4-20mA proportional electrical output signal and for Channel 2 to be 0-5000PSI for 4-20mA. Note: the pressure sensor with the smaller measurement range is selected for the maximum possible system pressure as both pressure sensors have only a single mechanical process connection. The use of identical pressure ranges or a reverse scaling of the second signal to 20-4 mA is also possible.

A random hardware or software error of the safety-critical load monitoring system can be detected by monitoring the dual pressure sensors in the control system, and the actuator can be set to a safe condition if a hazardous condition is detected. Therefore, the probability of a dangerous failure is substantially reduced. This solution increases the process security and is cost effective since the mechanical and electrical installation costs are clearly lower than installing two individual pressure sensors.

TÜV certified Safety
The MHS-1 is designed to meet both the ISO 13849-1 (performance level - PL) and IEC 61508 (Safety Integrity Level - SIL) safety standards. It is certified by TÜV, a recognized and independent test laboratory. The requirements of the new European Machinery Directive (2006/42/EC), which becomes mandatory in Europe at the end of 2009, have already been met.

Safety-related characteristics available from WIKA for the pressure sensor include, “Mean Time to Failure” – (MTTF values). A complete safety-related control system can be evaluated by the user quickly and easily. The MHS-1 achieves the Safety Integrity Level SIL 2 in accordance with IEC 61508 and safety category 3 Performance Level d in accordance with ISO 13849-1.

Robust Design for demanding environmental conditions
WIKA‘s hermetically-sealed, welded, dry thin-film measuring cell with its sputtered Wheatstone bridge offers long-term stability in applications with high dynamic load changes. Thin-film sensors feature excellent resistance to pressure spikes. In addition, an arc eroded pressure port and cavitation damping system integral to the process connection minimizes the possibility of pressure sensor failure. The electronics are designed for harsh EMC conditions up to 100 V/m. The robust stainless steel housing and the IP 69K high pressure steam washdown rated electrical connection are designed and extensively tested for the extreme conditions often encountered in mobile hydraulics.

Customer-specific Designs
WIKA has various safety-evaluated designs available for a variety of applications both with and without sensory overload detection. WIKA’s flexible instrument assembly and modern production system supports custom designs.

See the complete line of WIKA electronic pressure measurement products